Trouble shooting
guide
We collected the most common problems and their possible solutions in the trouble shooting guide below. If you are not able to find a solution to your problem in this guide, please do not hesitate to contact us.
GENERAL
INFORMATION
Generally higher voltage leads to a higher signal; however, this doesn’t necessarily translate to a higher S/N. A higher voltage also leads to more reactive species formed (N2+).
Recommended settings: 1300 – 1800 V. Do not set higher than 2000 V.
Frequency is the oscillation of the plasma turning off and on, therefore higher frequency usually increases the amount of reactive species which in hand usually leads to a higher signal and partially higher S/N.
Recommended settings: 15000 – 50000 Hz
With direct screening or GC/SPME module a recommended starting point is 15000 Hz.
With the LC module a recommended starting point is 50000 Hz.
Latest measurements with the LC module and high intake flow MS systems show that higher frequencies (e.g. 50kHz) can have a beneficial effect on RSDs and overall S/N.
Note: Please see paper of Gyr, Wolf, Franzke, Zenobi (Mechanistic Understanding Leads to Increased Ionization Efficiency and Selectivity in Dielectric Barrier Discharge Ionization Mass Spectrometry: A Case Study with Perfluorinated Compounds) for more details on fluorinated hydrocarbons for explanation and potential exploitation of these effects.
- The SICRIT ionization source is made of medical grade PEEK, (gold coated) copper and ceramic. It has long term temperature stability up to 260°C, and for short periods the PEEK can tolerate temperatures up to 320°C.
- The source cannot tolerate high concentrations of highly corrosive acids: Ex. HF, H2SO4, Aqua regia
- The source is not resistant against fluorine, chlorine, and bromine.
Do not use pure THF as it will cause swelling of the Peak.
The SICRIT® source in combination with modules (potentially direct screening as well) is partially facing problems with direct ESI swap on instruments that have very high flow inlet capillaries like Thermo Exploris 480 or TQ absolute. In combination the SICRIT® source and module, the SC-30 control unit may shut down and restart with error Low Voltage at HV-Output. To combat this issue, replace the inner electrode with a larger inner diameter (contact Plasmion).
Ion Source &
SC-30 Control Unit
If the source does not lock on the adapter make sure only a single o‑ring is installed on the ion source. Sometimes, o‑rings can get stuck on the adapter during removal of the source.
Potential Reason | Solution |
Source is clogged | Clean Source: Watch Cleaning Tutorial video. |
Core needs to be replaced | Replace Core: Watch Tutorial Video for Core Exchange. |
Module is not installed properly | Check installation:Operations and Installations Manual |
Module needs servicing | Contact Plasmion support team |
Issue with the Mass Spec | Contact MS Support |
Flow chart for problem solving in case of low signal:
Potential Reason | Solution |
Source is clogged | Clean Source: Watch Cleaning Tutorial video |
Core needs to be replaced | Replace Core: Watch Tutorial Video for Core Exchange |
SC-30 Control Unit needs servicing | Contact Plasmion support team |
Module needs servicing | Contact Plasmion support team |
Settings on the MS are not correct (e.g. tube lense voltage, gas flows, etc.) | Check Operations and Installations Manual |
Issue with the Mass Spec | Contact MS Support |
Flow chart for problem solving in case of no signal:
Potential Reason | Solution |
Core needs to be replaced | Replace Core: Watch Tutorial Video for Core Exchange |
Incorrect Injections | Make sure autosampler injects reproducibly (no bubbles, tight syringe, etc.) |
Problem with coupling to chromatography | Check injection port (septa and liners) and injection settings (speed, splitless time, injections depth, etc.) |
Potential Reason | Solution |
SC-30 Control issue is defect | Contact Plasmion support team |
Cable is defect | Contact Plasmion support team |
Plasma core is defect | Replace Core: Watch Tutorial Video for Core Exchange |
Ion source is defect | Contact Plasmion support team |
Flow chart for problem solving in case of high voltage error:
Potential Reason | Solution |
Electronics are overload, system shuts down and restarts | Do not vary the set values in short time intervals and wait until the actual value reached the target value |
Wrong settings | Recommended settings:
1. Voltage: 1300 – 1800 V 2. Frequency: 15000 Hz – 50000 Hz |
Instructions for firmware update:
- Contact Plasmion to receive an updated config file.
- To update the firmware, turn off the SC-30 control box, then remove the SD-card from the front of the box.
- Set the SD-card to write.
- Replace the files “FW_SC30_0XX.bin” and config.txt” with the new firmware and config files.
Note: Save the old files before replacing. - Restart the device. The LED ring should blink green during the rebooting process, indicating the firmware is being updated.
- By scrolling to the status tab, the new firmware version should be displayed.
GC/SPME
MODULE
Potential Reason | Solution |
Broken column or transfer column | Check the column inside the GC-/SPME-module |
Cold spot in GC-/SPME-module | Increase temperature of GC-/SPME-module |
Reduced column flow | Check for leakage at press fit connectors or inlet |
Potential Reason | Solution |
Cold spot in transfer line |
|
Cold spot in GC-/SPME-module |
|
Reduced column flow | Check for leakage at press fit connectors or inlet |
If the liner breaks during installation of the GC-SPME module, replace it and adjust the mounting to ensure an aligned connection between the Swagelok elbow and the SICRIT® source.
Potential Reason | Solution |
Excessive column bleed |
|
Roomair entering the system |
|
BREATH ANALYSIS
MODULE
Potential Reason | Solution |
Hydrocarbon residues in the nitrogen lines or bottles | Install additional gas filters in the nitrogen line |
Potential Reason | Solution |
Inverse suppression behaviour during exhalation | Install a t‑piece instead of the elbow and dilute the sample by a factor of 1:1 up to 3:1 with pure nitrogen. This should solve the suppression effect but will reduce the overall sensitivity. |
LC/SFC
MODULE
If you have issues with mounting the LC-Module onto the module adapter, make sure you have the mounting screws loosened to line up the module perfectly straight with the adapter. If it is still not going in, try slowly twisting the front part of the source with the adapter to the left while pushing it in the LC-Module.
Do not try to push with full force, especially if your MS has a glass inlet capillary. It might break!
Potential Reason | Solution |
Contamination in the LC System | Flush the LC according to the vendor manual |
Contamination in the adapter/sprayer | Clean adapter/sprayer |
Contamination in the LC-Module | Bake out the module 1–3 days at 500 °C unmounted from the MS, ideally with gas flow running through the module |
Room air entering the system | Check for leakage at press fit connectors of module |
Potential Reason | Solution |
Peak Tailing | Change pre-column and use fresh solvents. Change the column. Most of the time there is an issue with the LC column. |
Peak Splitting | Check all fittings (especially around the column) to prevent dead volume. Flush column backwards to get rid of particles and dirt. |
Peak Fronting | Try a lower concentration / higher dilution of the standards/samples. If possible, try to use the same solvent for diluting samples as for the mobile phase in the beginning of the LC run. |
Check for dead volume in the system especially at the fittings around the column and at the sprayer. If the separation is still bad or the peaks are very broad, unmount the sprayer from the module and hold it over a paper towel or in an empty beaker and start the pump to check if the spray is continuous and centered. In case of a blocked sprayer, clean it according to the cleaning tutorial. If separation is still an issue, there might be a problem with the column.
Potential Reason | Solution |
Sprayer blocked | Check and clean sprayer |
Potential Reason | Solution |
Capillary column blocked | Check and clean sprayer |